SS-300-NH3 Extractive Laser Analysis System adopts TDLAS Tunable diode laser absorption spectroscope to measure HF concentration. It adopts special heated sampling probe which is able to work in high-temperature and high dust conditions, and hot & wet sampling system with little gas loss caused by gas condensing. Compare in-situ type gas analyzer, the extractive system has the advantages of no requirement of installation conditions, easy maintenance, little pollution to mirror, suitable for dirty sampling point with high dust.
II. Applications field
Itis used to monitor NH3 after SNR or SCNR process in the waste incineration plant, power plant and so on.
III. Specifications
Monitoring place
Gas component
Measurement range
SNR or SCNR stack outlet
NH3
0-20ppm(can be customize)
1. Measurement principle:TDLAS
2. Sampling method:Hot & Wet extractive sampling
3. Sampling pipe material:SUS316
4. Heated tube diameter:ø8×6
5. Power supply: AC220V±10% 50/60Hz
6. Power consumption:Max 2000W+50W*heated tube length
7. Calibration period:each 6 months
8. Analogue output:4 - 20mA
9. Max load:4 - 20mA 550Ω
10. Alarm output:gas temperature alarm, gas low pressure, gas over pressure alarm.
IV. Features
1. Extractive sampling method
2. Easy calibration and calibration
3. Adopts TLDAS principle, little cross-sensitivity,
4. Laser detector with long lifespan
5. Optional sampling probe calibration function.
6. IP 65 protection class.
7. Adopts hot and wet sampling and gas pretreatment with temperature 190℃, preventing HCL gas loss by condensation.